Technical data
Table 7 - Electrical data iXH3030/T to pXH6000 (continued)
Characteristics iXH3030/T iXH3045H iXH3050H iXH4545HT iXH4550HT iXH6045H/T iXH6050H/T pXH4500 pXH6000 Units
Dry pump motor rating7.5111111111111 - - kW
Mechanical booster
motor rating
7.5 7.5 7.5 7.5 7.5 7.5 7.5 7.5 7.5 kW
Current rating
(200 - 230 V systems)
44 63 63 63 63 63 63 28 28 A
Current rating
(380 - 460 V systems)
23 32 32 32 32 32 32 14 14 A
Recommended branch
circuit fuse UL
(200 - 230 V systems)
55 80 80 80 80 80 80 35 35 A
Recommended branch
circuit fuse IEC
(200 - 230 V systems)
45 65 65 65 65 65 65 30 30 A
Recommended branch
circuit fuse UL
(380 - 460 V systems)
30 40 40 40 40 40 40 20 20 A
Recommended branch
circuit fuse IEC
(380 - 460 V systems)
25 35 35 35 35 35 35 15 15 A
Min cable size for 200 -
230 V systems (or
corresponding AWG
size)
10 (6) 25 (4) 25 (4) 25 (4) 25 (4) 25 (4) 25 (4) 6 (8)
*
*
The minimum geometric wire gauge for Han® K 4/4 is 6mm² and minimum AWG size is 8 AWG
6 (8)
*
mm
2
(AWG)
Min cable size for 380 -
460 V systems (or
corresponding AWG
size)
6 (8)
*
10 (6)
†
†
The minimum geometric wire gauge for Han® 100A module is 10mm² and minimum AWG size is 6 AWG
10 (6)
†
10 (6)
†
10 (6)
†
10 (6)
†
10 (6)
†
6 (8)
*
6 (8)
*
mm
2
(AWG)
Mains connector Han
®
K 4/4 Han
®
100 A
module
Han
®
100 A
module
Han
®
100 A
module
Han
®
100 A
module
Han
®
100A
module
Han
®
100A
module
Han
®
K 4/4 Han
®
K 4/4 -