SCU-1600 Control Unit for Turbomolecular Pump
March 10 19 Issue 1-a
3
INSTALLATION
3.5 Attaching the SCU-1600 to a rack
CAUTION
The SCU-1600 cannot be supported with only the 4 screws on the front panel. Always support it from
the bottom using a support angle. When installing the SCU-1600, DO NOT block the ventilation port
of the SCU-1600 by the support angle or other things. It will cause the SCU-1600 to overheat.
The dimensions of the SCU-1600 front panel conform to EIA standards. Therefore, this panel can be
attached to any type of commercially available rack. Attach the SCU-1600 as follows:
Note: For the dimensions of the front panel and positions of the screw holes for the rubber feet, refer
to Figure 9 and Figure 13.
• Attach the front panel to the rack using the 4 screws Figure 13 (1).
• Support the SCU-1600 from the bottom using a support angle Figure 13 (2) or a similar tool.
• When attaching the SCU-1600 to a movable rack, to protect the SCU-1600 during transport,
remove the rubber feet from the bottom and attach the SCU-1600 to the rack using the screw
holes for the rubber feet.
1
2
1. Front panel 4 securing screws 2. Mounting support
Figure 13 - Example of securing the SCU-1600