10.11 Refrigeration system failure (Ct)
The table below details the refrigeration system failure (Ct) parameter
DISPLAY
DESCRIPTION
Defines the maximum continuous runtime of the compressor without reaching the set
point (SP) temperature. If the set point (SP) temperature is not reached within this time,
the ems controller switches off the compressor and activates the refrigeration system (rSF)
alarm.
CONSIDERATOINS None
RANGE 00 to 100 hours
GLOBAL DEFAULT 72 hours
10.12 Display stability (d2)
The table below details the display stability (d2) parameter
DISPLAY
DESCRIPTION
Defines the rate of change of the displayed temperature.
Limiting the rate of change provides a dampening effect so as not to concern users should
the air temperature rise quickly due to a door opening.
Increasing the value for the display stability (d2) slows the rate of change of the displayed
temperature.
CONSIDERATOINS Use the global default value for normal operation
RANGE 1 to 254
GLOBAL DEFAULT 2