Technical data 23
3.4.2 Pressure Sensor Type 17002
Data
Value
Unit
External thread (internal model)
M12 x 1.5
Usable thread length (internal model)
approx. 10
mm
Measuring range
Overload capacity
0.9 7 bar abs.
10 bar abs.
The pressure sensor is available as both an external and
internal model. Further details can be found under
www.elster-instromet.com or chapter "
Assembly, Connection and Putting into Operation".
3.5 Temperature sensor
Data
Value
Unit
Measuring range
-30 … +60
°C
Measurement uncertainty
max. ± 0.1
%
Installation length
50
mm
3.6 Digital inputs
3.6.1 LF pulse and signalinputs
The maximum counting frequency of the digital inputs can be adjusted us-
ing the "enSuite" software. The limit values specified for frequency and du-
ration shall only apply if the so-called "software debounce" has been
switched off.
The software debounce is activated ex-works to suppress interference
pulses and therefore restrict safe readings to 2 Hz.
If the software debounce is parameterized to a frequency
higher than 2 Hz, then it may lead to counting errors
caused by electromagnetic interference under certain cir-
cumstances.