EasyManua.ls Logo

ELTRA ELEMENTRAC CS-i - Technical Specifications; Analyzer Specifications

ELTRA ELEMENTRAC CS-i
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data
29
7 Technical data
7.1 Technical data for the analyzer
The following table lists the technical data for the analyzer:
Definition
Description
Data
Dimensions
Width
52 cm
Depth
75 cm
Height
84 cm
Weight
Approx. 150 kg
Electrical data
Power supply
230 VAC ±10 %
Frequency
50/60 Hz
Power consumption
16 A
Mains connection
Safety plug
(Type F, CEE 7/4, IEC TR 60083)
Safety
Protection class
I
Overvoltage category
II
Degree of contamination
2
Type of environment
Indoors
Max. altitude
2000 m
Ambient temperature
+5 to +35°C
Ambient humidity
< 80 % at 31 °C
non-condensing
Type of protection
IP20
Carrier gas
Type
Oxygen
Purity
Min. 99.5 %
Pressure
2 to 4 bar (30 to 60 psi)
Connection
Polyamide hose 6/4 mm
(Adapter with G¼" female thread, included
with delivery)
Compressed air
Type
Oil-free
Pressure
5 to 6 bar (75 to 90 psi)
Connection
Polyamide hose 4/2 mm
(Adapter with G¼" female thread, included
with delivery)
Volume
Maximum noise level
55 db(A)
Tab. 1: Technical data for the analyzer