Maintenance
57
10.1.9 Replacing the O-rings
CAUTION
C29.0076
Risk of burns
Hot machine parts
Some parts of the machine can get very hot and cause burns when
touched without wearing protective gloves.
Wear heat-resistant protective gloves when working with hot
machine parts.
Never touch hot parts of the machine without suitable protective
gloves.
CAUTION
C30.0000
Risk of cuts
Glass splinters
Sample vials and reagent tubes are made of glass and can break. Glass
splinters can cause cuts.
Inspect reagent tubes and sample vials for damage prior to use.
Wear protective gloves and goggles when handling reagent tubes
and sample vials.
Replace damaged reagent tubes and sample vials.
Do not touch glass splinters with unprotected hands.
10.1.9.1 Replacing the O-ring at the top of the combustion tube
Replace the top O-ring on the combustion tube of the analyzer as follows:
1. Turn the main switch on the analyzer to Position 0 and pull out the power plug.
2. Wait until the analyzer has cooled down sufficiently.
3. Remove the furnace cleaning system, see Replacing the furnace cleaning brush and
thermal shield.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ELTRA ELEMENTRAC CS-i and is the answer not in the manual?

ELTRA ELEMENTRAC CS-i Specifications

General IconGeneral
BrandELTRA
ModelELEMENTRAC CS-i
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals