4. The meter will have to restart to install the firmware. Daniel MeterLink
will prompt
you with a message that it must disconnect from the meter. Once the firmware
upgrade is complete you will be able to reconnect to the meter with Daniel
MeterLink
.
5. Daniel MeterLink
prompts you with a message that it must disconnect from the
meter. Once the firmware upgrade is complete you will be able to reconnect to the
meter with Daniel MeterLink
.
Important
When the meter restarts, it takes about two minutes before you will be able to reconnect
depending on the firmware upgrade being performed. If the database does need to be re-
initialized, it could take up to five minutes.
6. Reconnect to the meter and repeat the Program Download process after the
Program Download. If all the program components are successfully updated, they
will show to be the same date and version as the Currently Installed Versions and the
Download button will be disabled. If one or more components are still not updated,
click Download to continue the upgrade process. Repeat the process to ensure all
components are updated.
NOTICE
The meters program may be lost if the meter loses power in the middle of a program
download.
If this occurs, perform the Backup Upgrade procedure to attempt a connection to
download the program again.
Error recovery for program download failure
If the Program Download failed during firmware upgrade, Daniel MeterLink can no longer
connect to the meter or a message Attempt FTP-only connection…. displays when trying
to connect perform a Backup Upgrade.
It is possible for the meter to lose its programming if the meter power fails during a
firmware upgrade process. If this occurs, use the following special procedure, Backup
Upgrade, to connect to the meter and download the program again.
Procedure
1. In Daniel MeterLink
, go to the File pull-down menu and select Program Settings.
The Program Settings dialog displays. Enable the Allow FTP-only connection
checkbox and click OK to close Program Settings.
a. For Serial Port Connections: Make sure you are connected to Port A. You may
need to adjust your File|Meter Directory settings for the connection so that they
match the port default settings. Port A will default to 19200 baud with a modbus
address of 32.