Configuration and Use Manual 83
Optional Configuration
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Some alarms can be reclassified. For example:
• The default severity level for Alarm A20 (calibration factors unentered) is
Fault, but you can
reconfigure it to either
Informational or Ignore.
• The default severity level for Alarm A102 (drive over-range) is
Informational, but you can
reconfigure it to either
Ignore or Fault.
For a list of all status alarms and default severity levels, see Table 8-7. (For more information on
status alarms, including possible causes and troubleshooting suggestions, see Table 11-4.)
To configure alarm severity:
• Using ProLink II, see Figure C-3.
• Using the Communicator, see Figure C-5.
Note: This functionality is not available via the display menus.
Table 8-7 Status alarms and severity levels
Alarm code
Communicator message
Default
severity Configurable
Affected by
fault timeoutProLink II message
A001 EEprom Checksum Error (Core Processor) Fault No No
(E)EPROM Checksum Error (CP)
A002 RAM Test Error (Core Processor) Fault No No
RAM Error (CP)
A003 Sensor Not Responding (No Tube Interrupt) Fault Yes Yes
Sensor Failure
A004 Temperature sensor out of range Fault No Yes
Temperature Sensor Failure
A005 Input Over-Range Fault Yes Yes
Input Overrange
A006 Transmitter Not Characterized Fault Yes No
Not Configured
A008 Density Outside Limits Fault Yes Yes
Density Overrange
A009 Transmitter Initializing/Warming Up Fault Yes No
Transmitter Initializing/Warming Up
A010 Calibration Failure Fault No No
Calibration Failure
A011 Excess Calibration Correction, Zero too Low Fault Yes No
Zero Too Low
A012 Excess Calibration Correction, Zero too High Fault Yes No
Zero Too High
A013 Process too Noisy to Perform Auto Zero Fault Yes No
Zero Too Noisy
A014 Transmitter Failed Fault No No
Transmitter Failed