EasyManua.ls Logo

Endress+Hauser Analytik Jena PlasmaQuant MS - Instrument Automated Setup; Torch alignment

Endress+Hauser Analytik Jena PlasmaQuant MS
139 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PlasmaQuant MS product family Optimizing hardware
59
6.1 Instrument Automated Setup
The ASpect MS can automatically perform plasma alignment, resolution and trim and
mass calibration routines.
All routines use the same default System Setup-worksheet: System Setup.msws. They
require the 1 µg/L (1 ppb) tuning solution.
Some customers use the instrument automated setup as a daily start-up of the
PlasmaQuant MS because you can add to the routine that the plasma is automatically
ignited and the system is warmed up.
1. Click the arrow next to the I
NSTRUMENT icon on the Main Toolbar and select
I
NSTRUMENT AUTO-SETUP to open the INSTRUMENT AUTOMATED SETUP window.
Alternatively select W
INDOW INSTRUMENT AUTO-SETUP.
2. Select the setup tasks you would like to perform by selecting the appropriate check
boxes in the T
ASK SCHEDULE / NAME column.
3. If desired, you can modify the task frequency and autosampler settings by
highlighting a task and clicking the [E
DIT] button.
4. Select the O
PEN REPORT AT COMPLETION check box to preview and print a report at
the end of the setup.
5. Click the [S
TART] button to start the initialization process. The area at the top of
the window will display the current task, and the estimated time left.
If you need to stop the set-up process, click the [C
ANCEL] button.
If all selected tasks passed, the instrument is ready to analyze your samples and you
can close the I
NSTRUMENT AUTOMATED SETUP window.
6.2 Torch alignment
It is important to ensure that the torch axis is perfectly aligned on the axis of the
sampler cone orifice for optimal signal efficiency. This should be done every time a
torch is installed or disturbed (e.g., moved aside to change/clean the cones).
It is a common practice to do this following routine once a week. It only takes about
five minutes to complete.
When done automatically, the best XY position found is stored by the instrument until
the next adjustment. The peak is found by summing all of the enabled isotopes
together and then selecting the location of the maximum of this signal. This means
that sometimes the selected peak may not be coincident with the maxima of the
largest signal scan present.

Table of Contents

Related product manuals