PlasmaQuant MS product family Autosampler ASPQ 3300
119
11 Autosampler ASPQ 3300
11.1 Function and setup
Fig. 62 Autosampler
The autosampler allows fully automated routine analysis. It can be equipped with
three sample racks and two racks each with 6 or 11 special samples, e.g., standards
and QC solutions. The following sample racks are available:
Rack/number of samples Cup volume (mL)
6
11 (special samples)
50
15
21 50
24 30
40 20
60 14
90 7
The rinse station is installed on the autosampler. A hose pump delivers the rinse
solution from the wash bottle into the wash cup this action cleans the dipped
cannula inside and outside. Excess rinse solution is pumped into the waste container
during rinsing. The rinse solution for rinsing between samples is also taken from the
wash cup. A mains connection supplies operating voltage to the autosampler.