PlasmaQuant MS product family Installation and commissioning
47
4.3.4 Installation of the transfer tube
There are two different transfer tubes:
Standard transfer tube
Transfer tube for Aerosol Dilution (optional)
Fig. 20 Transfer tube for Aerosol Dilution (left), standard transfer tube (right)
The transfer tube for Aerosol Dilution has an extra connection for sheath gas. The
following instruction shows the installation of the transfer tube for Aerosol Dilution.
WARNING
Risk of eye and skin injuries due to UV and electromagnetic radiation!
If you are using a non
-standard sample introduction configuration, there may be a
danger of direct exposure to UV or electromagnetic radiation from the opening in the
side wall of the torch compartment.
A UV and EMR protective cover is supplied with your PlasmaQuant
MS instrument for
use with non
-standard configurations.
1. For Aerosol Dilution: fit the tubing and
gas connector to the transfer tube (if not
already fitted).
2.
Swing the mounting arm of the spray
chamber outward to allow access to the
opening in the side wall of the torch
compartment.
3. Open the torch compartment lid to allow
access to the lever arm of the transfer
tube protection mechanism.