Detector calibration PlasmaQuant MS product family
92
Fig. 51 Detector calibration tab
6. When the calibration is complete, a table is displayed.
Fig. 52 Detector attenuation correction factors viewer
7. Review the DETECTOR ATTENUATION CORRECTION FACTORS VIEWER table and verify both
attenuation factors (Medium and High) are calibrated for all isotopes of interest.
Also verify that a good RSD (<3 %) is obtained for the calibrated attenuation
factors.