Method development PlasmaQuant MS product family
72
Fig. 39 Scan settings
3. Select the SCAN MODE. For details on each SCAN MODE, see the software help.
4. Select the scan
SPACING, or the distance between measurement points on a single
mass peak.
Typical methods use P
EAK HOPPING scan mode, with the POINTS PER PEAK =1 in the
A
CQUISITION settings. In this case, the SPACING selection is irrelevant. See the
software help for more details about when to select different scan modes.
5. Select the
ACQUISITION MODE. If TIME RESOLVED is selected, type a SAMPLING TIME.
6. Type values for each
ACQUISITION parameter: POINTS PER PEAK, SCANS/REPLICATE, and
R
EPLICATES/SAMPLE. Acquisition time is calculated based on the parameters entered.
7. Set the detector
ATTENUATION CORRECTION. To apply the detector attenuation factors
from a template, click [B
ROWSE] and select an existing template file.
7.7 Configure sampling settings
Sample introduction parameters such as spray chamber temperature, can be set, as
well as the uptake delay or rinse time of the autosampler.
Typically, the uptake and rinse delay times of 60 s are sufficient for most applications.