Method optimization PlasmaQuant MS product family
80
8.3.2 Run automatic optimization
Fig. 44 Selecting an optimization routine
1. Select an optimization routine from WINDOWS OPTIMISE ION OPITCS.
2. Verify that the selected optimization routine is appropriate. There are routines for
normal and high sensitivity and for the iCRC mode. The auto optimization window
appears after selecting an optimization routine.
3. Select S
TART from the OPTIMIZE menu or press [SHIFT+F8].