Maintenance and care PlasmaQuant MS product family
118
8. Close the suction port with the blind flange.
9.
Connect the pump with the mains.
Start the pump and run it for about 30 seconds.
Stop the pump and disconnect the mains power connection.
12.
Drain the pump fluid. Re-fit the outer drain cap.
13. Open the filler cap just above the view port.
14.
Add fluid until it is halfway up the window. Do not fill the
pump over the mark MAX. The in
ternal exhaust filter would
stand in the fluid and it would thus loose its properties.
Close the filler cap by hand.
Attach the vacuum tubing to the suction port.
Connect the pump with the power cable to the power
connection on the rear side of the PlasmaQuant MS (12 in
Fig. 17 p. 39).
Switch on the power switch of the pump. The pump is
controlled from the PlasmaQuant MS. The pump will only
run when the PlasmaQuant MS is put into operation.