Description of device parameters Cerabar PMP21 IO-Link
56 Endress+Hauser
Description To prevent switching on and switching off at values around the switch point, you can
configure a delay for the specific points within a range of 0 to 50 s with a resolution of 2
decimal places.
If the measured value leaves the switching range during the configured delay time, the
delay time starts again.
Selection 0.00 to 50.00 s
Factory setting 0 s
Switchback delay time, output 1 (dR1)
Navigation Parameter → Signal Switching Channels 1.1 → Switchback delay time, output 1 (dR1)
Description To prevent switching on and switching off at values around the switch-back point, you can
configure a delay for the specific points within a range of 0 to 50 s with a resolution of 2
decimal places.
If the measured value leaves the switching range during the configured delay time, the
delay time starts again.
Selection 0.00 to 50.00 s
Factory setting 0 s
Switching Signal Channel 1.2
SSC1.2 Param. SP1
Navigation Parameter → Signal Switching Channels 1.2 → SSC1.2 Param. SP1
Description Switch point 1 of switching signal SSC1.2 for pressure
Selection No selection. The user is free to edit the values.
SSC1.2 Param. SP2
Navigation Parameter → Signal Switching Channels 1.2 → SSC1.2 Param. SP2
Description Switch point 2 of switching signal SSC1.2 for pressure
Selection No selection. The user is free to edit the values.
SSC1.2 Config. Logic
Navigation Parameter → Signal Switching Channels 1.2 → SSC1.2 Config. Logic
Description Logic for inverting the switching signal SSC1.2 for pressure