Diagnostics and troubleshooting Deltabar PMD75B HART
54 Endress+Hauser
11 Diagnostics and troubleshooting
11.1 General troubleshooting
11.1.1 General errors
Device does not respond
Possible cause: Supply voltage does not match the specification on the nameplate
Remedial action: Apply the correct voltage
Possible cause: The polarity of the supply voltage is wrong
Remedial action: Correct the polarity
Possible cause: The cables do not contact the terminals properly
Remedial action: Check the electrical contact between cables and correct if necessary
Values not visible on the display
Possible cause: Contrast setting is too weak or too strong
Remedial action: Increase the contrast setting (make screen darker) by pressing  and
 simultaneously
Remedial action: Decrease the contrast setting (make screen brighter) by pressing  and
 simultaneously
Possible cause: The plug of the display cable is not connected correctly
Remedial action: Connect the plug correctly
Possible cause: Display is defective
Remedial action: Replace the display
No backlighting on the display
Possible cause: The electric current and voltage are not correct.
Remedial action: Adapt the current and voltage to the specified values.
"Communication error" is indicated on the display when the device is started or the
display is connected
Possible cause: Electromagnetic interference influence
Remedial action: Check grounding of the device
Possible cause: Defective cable connection or display plug
Remedial action: Replace the display
Output current <3.6 mA
Possible cause: Signal cable is not wired correctly
Remedial action: Check wiring
Possible cause: Electronics module is defective
Remedial action: Replace electronics
HART communication not working
Possible cause: Communication resistor missing or incorrectly installed
Remedial action: Install the communication resistor (250 Ω) correctly
Possible cause: Commubox is connected incorrectly
Remedial action: Connect Commubox correctly
Possible cause: Commubox is not set to "HART"
Remedial action: Set selector switch of Commubox to "HART"
Communication via CDI interface not working
Possible cause: Wrong setting of the COM port on the computer
Remedial action: Check the setting of the COM port on the computer and correct it if
necessary
Device measures incorrectly
Possible cause: Parametrization error
Remedial action: Check and correct the parameter configuration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Deltabar PMD75B HART and is the answer not in the manual?

Endress+Hauser Deltabar PMD75B HART Specifications

General IconGeneral
BrandEndress+Hauser
ModelDeltabar PMD75B HART
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals