Diagnostics and troubleshooting Deltabar PMD75B HART
58 Endress+Hauser
11.2 Diagnostic information on local display
11.2.1 Diagnostic message
Measured value display and diagnostic message in alarm condition
Faults detected by the self-monitoring system of the measuring device are displayed as a
diagnostic message in alternation with the measured value display.
A0043759
1 Status signal
2 Symbols for Bluetooth, device locked (lock symbol), communication (2 arrows) and square root extraction
(root symbol)
3 Status symbol with diagnostic event
1
4
3
X X X X X X XX X X X X X X X XX X
2
S
S
XX
20.50
X
i
S801
A0043103
1 Status signal
2 Symbols for Bluetooth, device locked (lock symbol), communication (2 arrows) and square root extraction
(root symbol)
3 Status symbol with diagnostic event
4 Event text
Status signals
F
"Failure (F)" option
A device error has occurred. The measured value is no longer valid.
C
"Function check (C)" option
The device is in the service mode (e.g. during a simulation).