About this document Deltabar PMD75B HART
8 Endress+Hauser
Example:
• Sensor: 10 bar (150 psi)
• Upper range limit (URL) = 10 bar (150 psi)
• Calibrated/adjusted span: 0 to 5 bar (0 to 75 psi)
• Lower range value (LRV) = 0 bar (0 psi)
• Upper range value (URV) = 5 bar (75 psi)
TD =
URL
|URV - LRV|
In this example, the TD is 2:1. This span is based on the zero point.
1.5 Documentation
All available documents can be downloaded using:
• the serial number of the device (see cover page for description) or
• the data matrix code of the device (see cover page for description) or
• the "Downloads" area of the website www.endress.com
1.5.1 Supplementary device-dependent documentation
Additional documents are supplied depending on the device version ordered: Always
comply strictly with the instructions in the supplementary documentation. The
supplementary documentation is an integral part of the device documentation.
1.6 Registered trademarks
HART®
Registered trademark of the FieldComm Group, Austin, Texas, USA
Bluetooth®
The Bluetooth® wordmark and logos are registered trademarks of Bluetooth SIG, Inc. and
any use of these trademarks by Endress+Hauser is licensed. Other trademarks and trade
names are those of their respective owners.
Apple®
Apple, the Apple logo, iPhone, and iPod touch are trademarks of Apple Inc., registered in
the U.S. and other countries. App Store is a service mark of Apple Inc.
Android®
Android, Google Play and the Google Play logo are trademarks of Google Inc.