Correction of measuring
range
The device must have been calibrated
and the echoes suppressed before the
measuring range can be corrected.
The new empty distance "E" may not
exceed that used for the calibration.
Otherwise a new calibration is
necessary, see page 19.
# Keys Significance
1 If appropriate calibration and echo
suppression, p. 19
2 Bring level in tank to "empty",
Wait 2 minutes
3 Empty calibration (4 mA)
= zero
4 Bring level in tank to "full",
Wait 2 minutes
5 Full calibration (20 mA)
= span
After correction of measuring range:
"Empty" level E 4 mA
"Full" level F 20 mA
Lock keys
In order to avoid accidental or
unauthorised changing of the settings, the
keys can be locked for new entries.
# Keys Significance
1 Lock operation:
2 Unlock operation:
After the keys have been locked:
All entries, including those via the
operating matrix, are blocked
(VU 330, DXR 275, Commuwin II).
The contents of the matrix fields can be
read, however.
Entries can only be unlocked by using
the appropriate key combination, see
above.
Note!
After the keys have been used, screw
on the housing cover.
+
+
__
+
+
__
D
F
E
100%
20mA
0%
4mA
BA171Y36
empty distance "E"
of calibration (old)
+
+
BA171Y35
Note!
Warning!
Chapter 5 On-site calibration without display VU 330 Micropilot FMR 231
22 Endress+Hauser