EasyManua.ls Logo

Endress+Hauser FMR 231 - 7.3 Fault analysis

Endress+Hauser FMR 231
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.3 Fault analysis
Table 7.2 lists the most common measuring errors with possible remedies. If the first
measure is successful, the remaining steps are not required.
Fault Analogue output Possible reason Remedy
Measured value incorrect
Distancw D (V0H8)
correct?
Stilling well?
Check microwave factor,
see basic calibration, page 23
False echo, see below
Offset (V3H6) > 0?
Set to zero
Calibration /
Linearisation
correct?
Check E (V0H1) and F (V0H2),
if necessary re-enter application
parameter V0H3, page 35
or. simulation, page 39
4/20 mA settings
correct?
Re-enter V0H5 and V0H6, page 28
Call Service!
Measured value remains
constant on emptying
(No warning E641)
Measured value jumps to
higher value when tank almost
empty
Measured value jumps
sporadically with constant
level but waves on surface
False echoes from
fittings/nozzle
False echo suppression
see page 24
Optimise alignment,
see page 36
Build-up on or near
the antenna causing
weak echoes
Activate window suppression,
see page 38
Carefully clean antenna
Signal weakened -
by waves false
echo sometimes
stronger
Select application parameter 1/2
(V0H3), page 22, 23
Activate window suppression,
page 38
Increase output damping, page 28
Active false echo suppression,
page 24
Optimise alignment, page 36 or .
check mounting position,
page 9 - 11
if possible, select better position
Call Service!
20 mA
F m/ft
E m/ft
D m/ft (V0H8)
4mA t
expected
happened
no
yes
no
yes
no
yes
20 mA
4mA t
expected
happened
20 mA
4mA t
expected
happened
20 mA
4mA t
happened
expected
yes
no
yes
no
yes
Micropilot FMR 231 Chapter 7 Trouble-Shooting
Endress+Hauser 37

Related product manuals