Commissioning Proline Promag P 500 HART
124 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry
Factory setting
Separator A local display is provided. Select decimal separator for
displaying numerical values.
. (point)
, (comma)
. (point)
Backlight One of the following
conditions is met:
Order code for "Display;
operation", option F "4-line,
illum.; touch control"
Order code for "Display;
operation", option G "4-line,
illum.; touch control
+WLAN"
Switch the local display
backlight on and off.
Disable
Enable
Enable
* Visibility depends on order options or device settings
10.5.5 Performing electrode cleaning
The Electrode cleaning circuit submenu contains parameters that must be configured for
the configuration of electrode cleaning.
The submenu is only available if the device was ordered with electrode cleaning.
Navigation
"Setup" menu → Advanced setup → Electrode cleaning circuit
Electrode cleaning circuit
Electrode cleaning circuit
→  124
ECC duration
→  124
ECC recovery time
→  124
ECC cleaning cycle
→  125
ECC Polarity
→  125
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Electrode cleaning circuit For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Enable the cyclic electrode
cleaning circuit.
Off
On
Off
ECC duration For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Enter the duration of electrode
cleaning in seconds.
0.01 to 30 s 2 s
ECC recovery time For the following order code:
"Application package", option
EC "ECC electrode cleaning"
Define recovery time after
electrode cleaning. During this
time the current output values
will be held at last valid value.
1 to 600 s 60 s