Diagnostics and troubleshooting Proline Promag P 500 HART
148 Endress+Hauser
Error Possible causes Solution
Network connection not present or
unstable
WLAN network is weak. Operating device is outside of
reception range: Check network
status on operating device.
To improve network
performance, use an external
WLAN antenna.
Parallel WLAN and Ethernet
communication
Check network settings.
Temporarily enable only the
WLAN as an interface.
Web browser frozen and operation
no longer possible
Data transfer active Wait until data transfer or current
action is finished.
Connection lost 1. Check cable connection and
power supply.
2. Refresh the Web browser and
restart if necessary.
Content of Web browser
incomplete or difficult to read
Not using optimum version of Web
server.
1. Use the correct Web browser
version .
2. Clear the Web browser cache and
restart the Web browser.
Unsuitable view settings. Change the font size/display ratio
of the Web browser.
No or incomplete display of
contents in the Web browser
JavaScript not enabled
JavaScript cannot be enabled
1. Enable JavaScript.
2. Enter http://XXX.XXX.X.XXX/
basic.html as the IP address.
Operation with FieldCare or
DeviceCare via CDI-RJ45 service
interface (port 8000)
Firewall of computer or network is
preventing communication
Depending on the settings of the
firewall used on the computer or in
the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
Flashing of firmware with FieldCare
or DeviceCare via CDI-RJ45 service
interface (via port 8000 or TFTP
ports)
Firewall of computer or network is
preventing communication
Depending on the settings of the
firewall used on the computer or in
the network, the firewall must be
adapted or disabled to allow
FieldCare/DeviceCare access.
12.2 Diagnostic information via light emitting diodes
12.2.1 Transmitter
Proline 500 – digital
Different LEDs in the transmitter provide information on the device status.