Operation Proline Promag P 500 HART
142 Endress+Hauser
11.5 Adapting the measuring device to the process
conditions
The following are available for this purpose:
Basic settings using theSetup menu (→  94)
Advanced settings using theAdvanced setup submenu (→  118)
11.6 Performing a totalizer reset
The totalizers are reset in the Operation submenu:
Control Totalizer
Reset all totalizers
Navigation
"Operation" menu → Totalizer handling
Totalizer handling
Control Totalizer 1 to n
→  142
Preset value 1 to n
→  142
Reset all totalizers
→  142
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry
Factory setting
Control Totalizer 1 to n One of the following options is
selected in the Assign process
variable parameter
(→  120) of the Totalizer
1 to n submenu:
Volume flow
Control totalizer value. Totalize
Reset + hold
Preset + hold
Reset + totalize
Preset + totalize
Hold
Totalize
Preset value 1 to n One of the following options is
selected in the Assign process
variable parameter
(→  120) of the Totalizer
1 to n submenu:
Volume flow
Specify start value for totalizer.
Dependency
The unit of the selected
process variable is
specified for the totalizer
in the Unit totalizer
parameter (→  120).
Signed floating-point
number
0 l
Reset all totalizers Reset all totalizers to 0 and
start.
Cancel
Reset + totalize
Cancel
11.6.1 Function scope of the "Control Totalizer" parameter
Options Description
Totalize The totalizer is started or continues running.
Reset + hold The totaling process is stopped and the totalizer is reset to 0.
Preset + hold The totaling process is stopped and the totalizer is set to its defined start value
from the Preset value parameter.
Reset + totalize The totalizer is reset to 0 and the totaling process is restarted.