Proline Promag P 500 HART Electrical connection
Endress+Hauser 45
7.2 Connecting the measuring device: Proline 500 -
digital
NOTICE
Limitation of electrical safety due to incorrect connection!
‣
Have electrical connection work carried out by appropriately trained specialists only.
‣
Observe applicable federal/national installation codes and regulations.
‣
Comply with local workplace safety regulations.
‣
Always connect the protective ground cable before connecting additional cables.
‣
For use in potentially explosive atmospheres, observe the information in the device-
specific Ex documentation.
7.2.1 Connecting the connecting cable
L
WARNING
Risk of damaging the electronic components!
‣
Connect the sensor and transmitter to the same potential equalization.
‣
Only connect the sensor to a transmitter with the same serial number.
‣
Ground the connection housing of the sensor via the external screw terminal.
Connecting cable terminal assignment
+ – AB
61 62 63 64
+ – AB
61 62 63 64
4
2
1
4
3
5
6
A0028198
1 Cable entry for cable on transmitter housing
2 Protective earth (PE)
3 Connecting cable ISEM communication
4 Grounding via ground connection; on device plug versions grounding is through the plug itself
5 Cable entry for cable or connection of device plug on sensor connection housing
6 Protective earth (PE)
Connecting the connecting cable to the sensor connection housing
Connection via terminals with order code for "Sensor connection housing":
– Option A "Aluminum, coated"→ 46
– Option L "Cast, stainless"→ 46
Connecting the connecting cable to the transmitter
The cable is connected to the transmitter via terminals → 47.