Commissioning Proline Promag P 500 HART
132 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Pulse output simulation Set and switch off the pulse
output simulation.
For Fixed value option:
Pulse width parameter
defines the pulse width
of the pulses output.
Off
Fixed value
Down-counting
value
Off
Pulse value In the Pulse output
simulation parameter, the
Down-counting value option
is selected.
Set and switch off the pulse
output simulation.
0 to 65 535 0
Device alarm simulation Switch the device alarm on and
off.
Off
On
Off
Diagnostic event category Select a diagnostic event
category.
Sensor
Electronics
Configuration
Process
Process
Diagnostic event simulation Select a diagnostic event to
simulate this event.
Off
Diagnostic event
picklist (depends
on the category
selected)
Off
Logging interval Define the logging interval tlog
for data logging. This value
defines the time interval
between the individual data
points in the memory.
1.0 to 3 600.0 s
* Visibility depends on order options or device settings
10.7 Protecting settings from unauthorized access
The following write protection options exist in order to protect the configuration of the
measuring device from unintentional modification:
Protect access to parameters via access code →  132
Protect access to local operation via key locking →  76
Protect access to measuring device via write protection switch →  134
10.7.1 Write protection via access code
The effects of the user-specific access code are as follows:
Via local operation, the parameters for the measuring device configuration are write-
protected and their values can no longer be changed.
Device access is protected via the Web browser, as are the parameters for the measuring
device configuration.
Device access is protected via FieldCare or DeviceCare (via CDI-RJ45 service interface), as
are the parameters for the measuring device configuration.
Defining the access code via local display
1. Navigate to the Define access code parameter (→  128).
2. Define a max. 16-digit character string comprising numbers, letters and special
characters as the access code.
3. Enter the access code again in the Confirm access code parameter (→  128) to
confirm the code.
The -symbol appears in front of all write-protected parameters.