Proline Promag P 500 HART Commissioning
Endress+Hauser 131
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
entry / User
interface
Factory setting
Assign simulation process variable – Select a process variable for
the simulation process that is
activated.
• Off
• Volume flow
• Mass flow
• Corrected volume
flow
• Flow velocity
• Conductivity
*
Off
Process variable value – Enter the simulation value for
the selected process variable.
Depends on the
process variable
selected
0
Status input simulation – Switch simulation of the status
input on and off.
• Off
• On
Off
Input signal level In the Status input simulation
parameter, the On option is
selected.
Select the signal level for the
simulation of the status input.
• High
• Low
High
Current input 1 to n simulation – Switch simulation of the
current input on and off.
• Off
• On
Off
Value current input 1 to n In the Current input 1 to n
simulation parameter, the On
option is selected.
Enter the current value for
simulation.
0 to 22.5 mA 0 mA
Current output 1 to n simulation – Switch the simulation of the
current output on and off.
• Off
• On
Off
Value current output 1 to n In the Current output 1 to n
simulation parameter, the On
option is selected.
Enter the current value for
simulation.
3.59 to 22.5 mA 3.59 mA
Frequency output simulation 1 to n In the Operating mode
parameter, the Frequency
option is selected.
Switch the simulation of the
frequency output on and off.
• Off
• On
Off
Frequency value 1 to n In the Frequency output
simulation 1 to n parameter,
the On option is selected.
Enter the frequency value for
the simulation.
0.0 to 12 500.0 Hz 0.0 Hz
Pulse output simulation 1 to n In the Operating mode
parameter, the Pulse option is
selected.
Set and switch off the pulse
output simulation.
For Fixed value option:
Pulse width parameter
(→ 104) defines the
pulse width of the pulses
output.
• Off
• Fixed value
• Down-counting
value
Off
Pulse value 1 to n In the Pulse output
simulation 1 to n parameter,
the Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65 535 0
Switch output simulation 1 to n In the Operating mode
parameter, the Switch option
is selected.
Switch the simulation of the
switch output on and off.
• Off
• On
Off
Switch status 1 to n – Select the status of the status
output for the simulation.
• Open
• Closed
Open
Relay output 1 to n simulation – Switch simulation of the relay
output on and off.
• Off
• On
Off
Switch status 1 to n The On option is selected in
the Switch output simulation
1 to n parameter parameter.
Select status of the relay
output for the simulation.
• Open
• Closed
Open