Commissioning Proline Promag P 500 HART
112 Endress+Hauser
10.4.10 Configuring empty pipe detection
The Empty pipe detection submenu contains parameters that must be configured for the
configuration of empty pipe detection.
Navigation
"Setup" menu → Empty pipe detection
‣
Empty pipe detection
Empty pipe detection
→ 112
New adjustment
→ 112
Progress
→ 112
Switch point empty pipe detection
→ 112
Response time empty pipe detection
→ 112
Parameter overview with brief description
Parameter Prerequisite Description Selection / User
interface / User
entry
Factory setting
Empty pipe detection – Switch empty pipe detection on
and off.
• Off
• On
Off
New adjustment The On option is selected in
the Empty pipe detection
parameter.
Select type of adjustment. • Cancel
• Empty pipe adjust
• Full pipe adjust
Cancel
Progress The On option is selected in
the Empty pipe detection
parameter.
Shows the progress. • Ok
• Busy
• Not ok
–
Switch point empty pipe detection The On option is selected in
the Empty pipe detection
parameter.
Enter hysteresis in %, below
this value the measuring tube
will detected as empty.
0 to 100 % 10 %
Response time empty pipe detection In the Empty pipe detection
parameter (→ 112), the On
option is selected.
Enter the time before
diagnostic message S862 ''Pipe
empty'' is displayed for empty
pipe detection.
0 to 100 s 1 s
10.4.11 Configuring the HART input
Navigation
"Setup" menu → HART input
‣
HART input
Capture mode
→ 113
Device ID
→ 113