Operation Liquiline CM44P
128 Endress+Hauser
Menu/Setup/Additional functions/Measuring range switch
Function Options Info
Current output
Range lower unit Depends on the Operating
mode
Units are only queried for Operating mode =
Conductivity . The other units are pre-defined
and cannot be modified.
• Conductivity
S/m, mS/cm, µS/cm, S/cm, µS/m, mS/m
• Concentration
%
• TDS
ppm
• Resistance
Ωcm
Range lower value
Range upper unit
Range upper value
Limit switches
Range lower unit Depends on the Operating
mode
Units are only queried for Operating mode =
Conductivity . The other units are pre-defined
and cannot be modified.
• Conductivity
S/m, mS/cm, µS/cm, S/cm, µS/m, mS/m
• Concentration
%
• TDS
ppm
• Resistance
Ωcm
Range lower value
Range upper unit
Range upper value
10.6.7 Diagnostic modules
Here you can configure a maximum of 8 individual diagnostic messages.
A diagnostic module has the following properties:
• The feed source can be configured like a binary output (relay, digital output).
• You can choose whether the diagnostic message should be output at a high level or low
level.
• You decide which error category (Namur class) should be assigned to the message.
• You can define a customized text to be output as the diagnostic message text.
In addition you can switch off the factory diagnostic code for limit switches. This allows
you to:
• Use the limit switch on a purely functional basis (without a message)
• Configure application-specific message texts
• Control diagnostic modules directly by a digital signal or via a limit switch output
(enables the use of switch-on/switch-off delay, for example).
Menu/Setup/Additional functions/Diagnostic modules
Function Options Info
Diagnostic module 1 (961) ...
8 (968)
Source of data Selection
• None
• Fieldbus signals
• Binary inputs
• Limit switches
Factory setting
None
Specify the input which should be the source of
data for the diagnostic message.
Measured value Selection
depend on Source of data
Factory setting
None
Specify the measured value that should trigger
the diagnostic message.
You can use different measured values
depending on the data source. → 100