Diagnostics and troubleshooting Liquiline CM44P
158 Endress+Hauser
No. Message Factory settings Sensor
type
Tests or remedial action
S
1)
D
2)
F
3)
950 Process
temperature
F On On C Concentration table (conductivity)
Process temperature below the lowest
value in the table
Extend table.
951 Process
temperature
F On On C Concentration table (conductivity)
Process temperature above the highest
value in the table
Extend table.
952 Conductivity low F On On C Concentration table (conductivity)
Process conductivity below the lowest
value in the table
Extend table.
953 Conductivity high F On On C Concentration table (conductivity)
Process conductivity above the highest
value in the table
Extend table.
954 Concentration low F On On C Concentration table (conductivity)
Process concentration below the lowest
value in the table
Extend table.
955 Concentration high F On On C Concentration table (conductivity)
Process concentration above the highest
value in the table
Extend table.
983 Sensor ISE check F On On I Electrode or membrane defective
1. Check or replace electrode.
2. Check or replace membrane cap.
984 Process temp. high S On On I Temperature outside specification
1. Check process temperature.
2. Check electrode system.
985 Sensor Interface F On On I Sensor interface error
1. Check connector.
2. Check or replace cable.
987 Calibration
required
M On On I, DI Electrode replacement
Calibrate sensor.
1) Status signal
2) Diagnostic message
3) Failure current