Endress+Hauser Liquiline CM44P - 16.10 Performance Characteristics

Endress+Hauser Liquiline CM44P
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquiline CM44P Technical data
Endress+Hauser 195
Cable entries
Identification of the cable entry on housing base Suitable gland
B, C, H, I, 1-8 M16x1.5 mm/NPT3/8"/G3/8
A, D, F, G M20x1.5 mm/NPT1/2"/G1/2
E -
M12x1.5 mm
H 5 6 7 8 I
GF
E
D
C
A
1 2 3 4 B
A0018025
Recommended assignment
1-8
A
B
C
D,F,G
H
I
E
Sensors 1-8
Power supply
RS485 In or M12 DP/RS485
Can be used freely
Current outputs and inputs, relays
Can be used freely
RS485 Out or M12 Ethernet
Do not use
Cable specification
Cable gland Permitted cable diameter
M16x1.5 mm 4 to 8 mm (0.16 to 0.32")
M12x1.5 mm 2 to 5 mm (0.08 to 0.20")
M20x1.5 mm 6 to 12 mm (0.24 to 0.48")
NPT3/8" 4 to 8 mm (0.16 to 0.32")
G3/8 4 to 8 mm (0.16 to 0.32")
NPT1/2" 6 to 12 mm (0.24 to 0.48")
G1/2 7 to 12 mm (0.28 to 0.48")
Cable glands mounted at the factory are tightened with 2 Nm.
Length of display cable provided (cabinet device only):
3 m (10 ft)
Maximum permitted length of a display cable (cabinet device only):
5 m (16.5 ft)
16.10 Performance characteristics
Response time Current outputs
t
90
= max. 500 ms for an increase from 0 to 20 mA
Current inputs
t
90
= max. 330 ms for an increase from 0 to 20 mA
Digital inputs and outputs
t
90
= max. 330 ms for an increase from low to high
Reference temperature 25 °C (77 °F)