Electrical connection Liquiline CM44P
32 Endress+Hauser
NOTICE
The device does not have a power switch!
Provide a protected circuit breaker in the vicinity of the device at the place of
installation.
The circuit breaker must be a switch or power switch, and must be labeled as the circuit
breaker for the device.
At the supply point, the power supply must be isolated from dangerous live cables by
double or reinforced insulation in the case of devices with a 24 V supply voltage.
6.2.1 Opening the housing
NOTICE
Pointed or sharp tools
If unsuitable tools are used, they can scratch the housing or damage the seal, and thus
have a negative impact on the leak-tightness of the housing!
Do not use pointed or sharp objects,e.g. knives, to open the housing.
Only use a suitable Phillips screwdriver.
1.
2.
3.
4.
 31 Releasing housing screws in a diagonally
opposite sequence with Phillips screwdriver
180°
 32 Opening display cover, max. opening angle
180˚ (depends on installation position)
1. Release the housing screws on a step-by-step basis. Start with any screw of your
choice. Then release the screw diagonally opposite this screw etc.
2. To close the housing: tighten the screws in a similar step-by-step, diagonally opposite
sequence.