Liquiline CM44P Electrical connection
Endress+Hauser 43
Example: Connecting the cleaning unit 71072583 for CAS40D
NOTICE
Power consumption too high for the Liquiline alarm relay!
Can cause irreparable damage to the base module
Connect the cleaning unit only to terminals of an additional module (AOR, 2R or 4R),
not to the alarm relay of the base module.
PE L1 N PE
T2.0A
230 V AC / 50 Hz
1
41
43
42
Liquiline/Sampler/
Analyzer:
AOR/2R/4R
A0028597
 61 Connecting the cleaning unit for CAS40D
Example: Connecting the Chemoclean CYR10 injector cleaning unit
41
43
42
Relay 1
41
43
42
Relay 2
1
2
1
2
NL PE
CM44x/CM44xR/
CM44P
CYR10
1
2
3
4
5
A0028598
 62 Connecting the CYR10 injector cleaning unit
1 External power supply
2 Cleaner to spray head
3 Container with cleaner
4 Motive water 2 to 12 bar (30 to 180 psi)
5 Backflow valve (to be provided by the customer)