2. Go to Menu/Setup/General settings/Extended setup/Data management/Load
setup .
A list of all the setups on the SD card is displayed.
An error message is displayed if there is no valid setup on the card.
3. Select the desired setup.
A warning is displayed:
The current parameters will be overwritten and the device will reboot.
Warning: Please note that cleaning and controller programs can be active.
Do you want to proceed?
4. Use OK to confirm or cancel.
If you select OK to confirm, the device restarts with the desired setup.
Exporting the setup
Exporting a setup offers the following advantages, among others:
• Export in XML format with a stylesheet for formatted display in an XML-compatible
application, such as . Microsoft Internet Explorer
• Importing the data (drag and drop the XML file into a browser window)
1. Insert the SD card into the controller card reader.
2. Go to Menu/Setup/General settings/Extended setup/Data management/Export
setup .
3. Name: Assign a file name.
4. Then select Export .
If you have already assigned the file name, you will be asked whether you want to
overwrite the existing setup.
5. Use OK to confirm or cancel and assign a new file name.
Your setup is saved on the SD card in the "Device" folder.
You cannot upload the exported setup to the device again. You must use the Save
setup . This is the only way you can save a setup to an SD card and reload it later on or
upload it to other devices.
Activation code
You require activation codes for:
• Additional functions, e.g. fieldbus communication
• Firmware upgrades
• Modifications, e.g. deactivation of fieldbus protocols
If your original device has activation codes, these can be found on the nameplate. The
corresponding device functions are activated at the factory. You only require the codes
if servicing the device or deactivating fieldbus protocols.
1. Enter activation code: Menu/Setup/General settings/Extended setup/Data
management/Activation code.
2. Confirm your entry.
The new hardware or software function is activated and can be configured.
Functions that are enabled by an activation code:
Function Activation code beginning with
Second Memosens input 062...
Deactivation of fieldbus when the 485 module is removed
1)
0B0...
Two current outputs (BASE2-E module only) 081...
Web server