EasyManua.ls Logo

Endress+Hauser Micropilot FMR62 - Power Supply and Wiring; Connection Diagrams and Voltage

Endress+Hauser Micropilot FMR62
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR62
Endress+Hauser 17
Power supply
Terminal assignment
Terminal assignment 2-wire: 4-20 mA HART
1
+
2
-
A
1
+
2
-
B
1
2
3
A0036498
 3 Terminal assignment 2-wire: 4-20 mA HART
A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection 4-20 mA HART passive: terminals 1 and 2, without integrated overvoltage protection
2 Connection 4-20 mA HART passive: terminals 1 and 2, with integrated overvoltage protection
3 Terminal for cable screen
Block diagram 2-wire: 4-20 mA HART
5
4
1
2
3
++
--
Y
I
6
1
2
A0036499
 4 Block diagram 2-wire: 4-20 mA HART
1 Active barrier with power supply (e.g. RN221N); observe terminal voltage
2 HART communication resistor (≥ 250 Ω); observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device; observe maximum load
5 Cable screen; observe cable specification
6 Measuring device

Table of Contents

Other manuals for Endress+Hauser Micropilot FMR62

Related product manuals