Installation in the bypass
1
1
A0042914
 17 Installation in the bypass
1 Tank connectors
Measurements can be performed through an open full bore ball valve without any problems.
Recommendations for the bypass
Metal (no enamel plastic or enamel liner).
Constant diameter.
Select an antenna that is as big as possible. Recommendation, use the 80mm/3" antenna.
Difference in diameter between antenna and inner diameter of the bypass as small as possible.
At transition points, e.g. when a ball valve is used or individual pipe segments are joined, any gaps
that occur should not exceed 1 mm (0.04 in).