Micropilot FMR62
38 Endress+Hauser
1. Pull out the display module with a gentle rotational movement.
2. Turn the display module to the desired position: max. 8 × 45 ° in each direction.
3. Feed the coiled cable into the gap between the housing and main electronics module and plug
the display module into the electronics compartment until it engages.
Closing the cover of the electronics compartment
A0021451
1. Screw down the cover of the electronics compartment.
2. Turn the securing clamp 90 ° clockwise and tighten the screw of the securing clamp of the
electronics compartment cover with 3 mm using the Allen key (2.5 Nm).
Post-installation check
Is the device undamaged (visual inspection)?
Does the device comply with the measuring point specifications?
• Process temperature
• Process pressure
• Ambient temperature range
• Measuring range
Are the measuring point identification and labeling correct (visual inspection)?
Is the device adequately protected against precipitation and direct sunlight?
Is the device adequately protected against impact?
Are all mounting and safety screws securely tightened?
Is the device properly secured?