Micropilot FMR62
Endress+Hauser 41
Vibration resistance
DIN EN 60068-2-64 IEC 60068-2-64 at 5 to 2 000 Hz: 1.5 (m/s
2
)
2
/Hz
A limited vibration resistance applies according
DIN EN 60068-2-64/IEC 60068-2-64 at 5 to 2 000 Hz of 0.39 (m/s
2
)
2
/Hz,
if the following order features are selected in the product structure at the same time:
• 040 ("Housing"): B ("GT18 two-chamber, 316L)
and
• 090 ("Seal"):
• A6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
or
• C2 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
or
• F6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
EN 61326 series and NAMUR Recommendation EMC (NE 21). For details, refer to the Declaration of
Conformity.
Download at www.endress.com
A normal device cable suffices if only the analog signal should be used. Use a shielded cable for
digital communication (HART/ PA/ FF).
Always use a shielded cable for the electronic version "2-wire, 4-20 mA HART + 4-20 mA analog".
Maximum measured error during EMC testing: < 0.5 % of the span. By way of derogation, for
devices with a plastic housing and see-through lid (integrated display SD02 or SD03) the measured
error can be up to 2 % of the span in the event of strong electromagnetic radiation in the 1 to 2 GHz
frequency range.