72
11.1.3 Error Table
No.
E001
E002
E003
E004
E005
E006
E007
E010
E011
E015
E016
E036
E037
E038
E039
E040
E041
E049
E050
E051
E052
Display
Internal communication error
Data ERROR in EEPROM
Invalid configuration
Configuration changed
Unknown card identifier
Checksum error in EEPROM
Transmitter defective
Temperature sensor defective
Temperature sensor 2 defective
Current loop 1 open
Current loop 2 open
Calibration range of meas. cell 1
exceeded
Below calibration range of meas.
cell 1
Calibration range of meas. cell 2
exceeded
Below calibration range of meas.
cell 2
TC table invalid
Substance table invalid
Adaptation factor range for meas.
cell 1 exceeded
Below adaptation factor range for
meas. cell 1
Adaptation factor range for meas.
cell 2 exceeded
Below adaptation factor range for
meas. cell 1
Remedy
Return the instrument to your
Endress+Hauser sales agency to be
repaired or request service assistance.
Check slot configuration
Set new configuration with set config
in Service / Factory settings menu
Check modules in slots
Perform checksum correction in
Special functions menu
Return instrument to your
Endress+Hauser sales agency for
repair or request service assistance
Check temperature measurement and
connections; check instrument and
measuring cable with temperature
simulator if necessary
Check connections, cables and any
connected instruments
Recalibrate measuring cell; check
measuring cell and connections;
check instrument and measuring
cable with temperature simulator
Correct temperature compensation
table; check temperature measurement
and connections; if necessary, check
instrument and measuring cable with
temperature simulator
Correct substance table
Continued on the next page...
Failure
Maintenance Required