73
No.
E055
E056
E057
E058
E059
E060
E061
E062
E063
E064
E065
E066
E067
E068
E073
E074
E075
E076
E077
E078
E080
E081
E142
E143
E144
E145
E148
E149
Display
Below display range of meas. value
Below display range of meas.
value 2
Display range of measured value
exceeded
Display range of measured value 2
exceeded
Below temperature range
Below temperature range 2
Temperature range exceeded
Temperature range 2 exceeded
Current limiting, 0/4 mA, output 1
Current limiting, 20 mA, output 1
Current limiting, 0/4 mA, output 2
Current limiting, 20 mA, output 2
Limit or setpoint 1 exceeded
Limit or setpoint 2 exceeded
Below TC range
Below TC range of channel 2
TC range exceeded
TC range of channel 2 exceeded
Temperature outside TC table range
Temperature 2 outside TC table
range
Range for current output 1 too small
Range for current output 2 too small
Break point outside range, current
output 1
Break point outside range, current
output 2
Range of current output 1 for
selected MR (= measuring range)
too small
Range of current output 2 for
selected MR (= measuring range)
too small
Break point outside range of current
output 1 MRX (= current measuring
range)
Break point outside range of current
output 2 MRX (= current measuring
range)
Remedy
Check measurement, control and
connection; if necessary, check
instrument and measuring cable with
simulator
Check the configuration in the Current
outputs menu; check measurement,
control and connections; if necessary
check instrument and measuring cable
with simulator
Check configuration in Relay contacts
menu; check measurement, control
and connections; if necessary, check
instrument and measuring cable with
simulator
Check configuration in Temperature
menu; check measurement, control
and connections; if necessary, check
instrument and measuring cable with
simulator
Check temperature compensation
table and correct if necessary, check
temperature measurement and
connections; if necessary, check
instrument and measuring cable with
temperature simulator
Increase range in Current outputs
menu
Correct configuration in Current
outputs menu
Increase range in Current outputs
menu
Correct configuration in Current
outputs menu
Disturbances
Warnings

Table of Contents