77
11.5.5 Special Functions
Function Selection
Optimization,
Selection of special functions Checksum correction,
Reset
For Optimization selection
Determination of temperature Determination of temperature coefficient Tc of a medium.
coefficient
Request sample 1 Immerse the measuring cell and temperature sensor in
a medium sample. The temperature of the sample
should be as close as possible to the reference
temperature used.
Measurement of Press the E key when both values are stable.
conductivity and
temperature
Request sample 2 Heat the sample by at least 10 Kelvin.
Immerse the measuring cell and temperature sensor
in the sample.
Measurement of Press the E key when both values are stable.
conductivity and
temperature
Display of temperature Note result for further use.
coefficient
Sensor preparation Hold sensor in air
Calibration runs Value is controlled for stability, next step is started
automatically
Air set information Displays OK/fail
Entry air set value 180.0 µS/cm to 180.0 µS/cm
channel 1
Entry air set value 180.0 µS/cm to 180.0 µS/cm
channel 2
For Checksum correction selection
Checksum correction In the case of checksum error E006, this function permits
an elimination of the error without resetting the instrument
to the factory settings.
Warning!
Check all settings when this function has
been used. Data loss is possible!
For Reset selection
Warm boot The instrument acts in the same way as when it is
switched off and then back on again.