77
11.5.5 Special Functions
Function Selection
Optimization,
Selection of special functions Checksum correction,
Reset
For Optimization selection
Determination of temperature Determination of temperature coefficient Tc of a medium.
coefficient
Request sample 1 Immerse the measuring cell and temperature sensor in
a medium sample. The temperature of the sample
should be as close as possible to the reference
temperature used.
Measurement of Press the E key when both values are stable.
conductivity and
temperature
Request sample 2 Heat the sample by at least 10 Kelvin.
Immerse the measuring cell and temperature sensor
in the sample.
Measurement of Press the E key when both values are stable.
conductivity and
temperature
Display of temperature Note result for further use.
coefficient
Sensor preparation Hold sensor in air
Calibration runs Value is controlled for stability, next step is started
automatically
Air set information Displays OK/fail
Entry air set value 180.0 µS/cm to 180.0 µS/cm
channel 1
Entry air set value 180.0 µS/cm to 180.0 µS/cm
channel 2
For Checksum correction selection
Checksum correction In the case of checksum error E006, this function permits
an elimination of the error without resetting the instrument
to the factory settings.
Warning!
Check all settings when this function has
been used. Data loss is possible!
For Reset selection
Warm boot The instrument acts in the same way as when it is
switched off and then back on again.

Table of Contents

Other manuals for Endress+Hauser Mycom CLM 152

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Mycom CLM 152 and is the answer not in the manual?

Endress+Hauser Mycom CLM 152 Specifications

General IconGeneral
BrandEndress+Hauser
ModelMycom CLM 152
CategoryTransmitter
LanguageEnglish

Related product manuals