OXY5500 Gas Analyzer Operating Instructions
Endress+Hauser 25
5.7.1 To enter editing mode
1. Click OK to enter editing mode.
2. Change the setting or value (one digit at a time) using the Up and Down Arrow buttons.
3. Click OK again to save editing changes.
5.7.2 To exit the editing mode
1. Click Menu to cancel and exit.
5.8 Analog output settings (Analogues) menu
From the MAIN MENU, select Analogues to change the analog output settings. Refer to Figure 24.
Figure 24. Main menu screen - Analogues selected
Before displaying the ANALOGUES screen, a message window displays requesting confirmation to abort the currently
running operation. Refer to Figure 25.
Figure 25. Message window - Stop measurements during configuration
Select Yes and stop the measurement to proceed with the Analog Output settings.
The ANALOGUES menu is divided into four screens: 4-20mA INTERFACE SETTINGS, 4-20mA VALUES,
CONCENTRATION ALARM RELAY (LS2), and 4-20mA CALIBRATION. Refer to Analog Output Settings (Analogues)
Menu Options → .
Use the Up and Down Arrow buttons to navigate between input fields.
5.8.1 To enter editing mode
1. Click OK to enter editing mode.
2. Change the setting or value (one digit at a time) using the Up and Down Arrow buttons.
3. Click OK again to save editing changes.