5.10 Measurement settings (Meas. settings) menu options
After selecting Meas. Settings from the MAIN MENU, the MEASUREMENT SETTINGS window displays. The analyzer
temperature compensation, pressure compensation, interval, and logging and data management options are
accessed from this screen.
5.10.1 Temperature compensation
On the MEASUREMENT SETTINGS screen, use the Navigation buttons to move to the Temperature box. Refer to
Figure 32.
Figure 32. Measurement settings screen - Temperature compensation
With Auto selected, the measurement temperature is determined by the RTD (Pt100) sensor.
Automatically measured temperature values can be displayed in °C, °F, or K.
5.10.2 Setting the temperature compensation
1. Change the settings to the desired unit of measurement in the lower right corner of the Temperature box. Figure
32 shows the temperature set at 22.0 °C.
OR
Select Manual if the temperature during measurement at the oxygen sensor is known and constant throughout
the measurement.
Manual setting is only required if the temperature probe is not functioning properly. Refer to Service →  before
using the Manual setting.
Temperature values can be input in °C, °F, or K, in a range from -99 °C to 199 °C. Values will automatically be
recalculated in the respective unit.
2. Switch to the desired temperature unit and change the temperature value in the input field to the measurement
temperature.