OXY5500 Gas Analyzer Operating Instructions
Endress+Hauser 49
5.14 Digitals menu options
Set RS-232, RS-485, and TCP/IP configurations from the Digitals button on the MAIN MENU.
5.14.1 RS-232 settings
Use this screen to set the baud rate of the RS-232 channel. Refer to Figure 61.
Figure 61. Digitals - RS-232 settings
The Baudrate for the RS-232 channel can be set to 9600, 19200, 38400, 57600, or 115200.
The ID, which is used in the Modbus communication, can be set to any value between 1 and 32.
Parity can be set to Even, Odd, or None.
Setting the Parity to “None” also sets the number of stop bits to two. Odd and Even settings will use one stop bit.
All settings are applied by clicking Save.
5.14.2 RS-485 settings
Use this screen to set the baud rate of the RS-485 channel. Refer to Figure 62.
Figure 62. Digitals - RS-485 settings
The Baudrate for the RS-485 channel can be set to 9600, 19200, 38400, 57600, or 115200.
The ID, which is used within the Modbus communication, can be set to any value between 1 and 32.
Parity can be set to Even, Odd, or None.
Setting the Parity to “None” also sets the number of stop bits to two. Odd and Even settings will use one stop bit.
All settings are applied by clicking Save.