5.14.3 TCP/IP settings
Use this screen to set TCP/IP. Refer to Figure 63.
Figure 63. Digitals - TCP/IP settings
• If DHCP is selected, the IP and Subnet Mask will be assigned by the DHCP server and therefore are not editable.
• If Static is selected, the IP and Subnet Mask must be entered manually. Contact your local Network
Administrator for assistance if input data needs to be confirmed.
• Port specifies the network port under which the Modbus application takes place. The default value for most
Modbus applications is 502.
• The ID, which is used with the Modbus communication, can be set to any value between 1 and 32.
All settings are applied by clicking Save.
5.15 Analog output settings (Analogues) menu options
From the MAIN MENU, click Analogues to access the 4-20mA INTERFACE SETTINGS, 4-20mA VALUES,
CONCENTRATION ALARM RELAY (LS2), and 4-20mA CALIBRATION screens.
5.15.1 4-20mA interface settings
The 4-20mA INTERFACE SETTINGS can be accessed through the ANALOGUES menu. When accessed, the following
screen displays. Refer to Figure 64.
Figure 64. Analogues - 4-20mA interface settings
The Output, Mode, and Error Trigger Level settings will be applied to the selected Port, which includes Port1, Port2,
or the Input.
The Output of Port1 or Port2 can be Oxygen or Temperature.
The Input is always Pressure and cannot be changed.