Operating Instructions OXY5500 Gas Analyzer
26 Endress+Hauser
5.8.2 To exit editing mode
1. Click Menu to cancel and exit editing mode.
All changes will be applied after the next measurement period.
5.9 Measurement menu options
Selecting Measurement from the MAIN MENU will open the SIMPLE screen. The DETAILS or GRAPH screens can be
selected from the SIMPLE screen.
5.9.1 Simple screen
This screen displays the oxygen and temperature values from the moment the measurement was started. Refer to
Figure 26.
Figure 26. Simple measurement screen
If the measurement temperature was set manually, the temperature value is already shown before starting the
measurement.
In manual mode, the temperature unit can be changed. Values ranging from -99 °C to 199 °C can be input into
the MEAS. SETTINGS window. Refer to Temperature compensation → .
If automatic temperature measurement is selected and the temperature sensor is not connected or not working
properly, the display will show an error message. Refer to Figure 27.
Figure 27. Temperature sensor error message
If no sensor is connected or is not connected properly, and the signal cannot be read when measurements are started,
an error message will be displayed in the status bar as shown in Figure 28.