OXY5500 Gas Analyzer Operating Instructions
Endress+Hauser 27
Figure 28. Error message - Sensor cannot be detected
The oxygen values are displayed in the following units:
For OP-3 sensor: %O2
For OP-6 sensor: %O2, ppmv
For OP-9 sensor: ppmv
1. Click the Up and Down Arrow buttons to change the oxygen unit on the display. The last measurement value
shown in the respective oxygen unit immediately. Choose one of the following options:
o Click the Right Arrow to display the detailed measurement screen. Refer to Details screen → .
o Click the Left Arrow to display the measurement graph. Refer to Graph screen → .
2. Click Menu to return to the MAIN MENU screen.
5.9.2 Details screen
The DETAILS screen gives additional information about measurement and measurement settings. Refer to Figure 29.
Figure 29. Details measurement screen
This screen is sectioned into boxes containing information on Oxygen, Temperature, Measurement Name, and
General.
Oxygen: This box displays the last measured value in the selected oxygen unit. It also shows the phase angle and
amplitude values. Change the oxygen unit by clicking the button.
Temperature: In this box, the current, last measured or manually set temperature value is displayed in the
selected temperature unit.
The temperature unit can be changed in Manual mode. Values ranging from -99 °C to 199 °C can be input into
the MEAS. SETTINGS window. Refer to Temperature compensation → .
Measurement Name: This box displays the selected measurement file in which all data is stored when logging is
turned on.
The measurement file can be changed in the MEAS. SETTINGS menu. Refer to Logging and data management →
.