Operating Instructions OXY5500 Gas Analyzer
66 Endress+Hauser
7 Appendix A: Specifications
Application Data
Target Components O2
Principle of Measurement Fluorescent Quenching
OP-9 OP-6 OP-3
Typical Measurement Ranges 0 to 200 ppmv (default)
0-10 to 10-1,000 ppmv
User setting
0 to 5%
0-1 to 0-5%
User setting
0-20%
0-10 to 0-20%
User setting
Lower Limit of Detection 0.5 ppmv 20 ppmv 300 ppmv
Accuracy at 20 to 25 °C ±5% of Reading ±3% of Reading ±2% of Reading
Repeatability ±1% of Reading
Measurement Update Time Programmable Sampling Rate (default 30 seconds)
Temperature Range
(Configurable)
1) 0 °C to 60 °C (0 °F to 140 °F)
2) -20 °C to 50 °C (-4 °F to 122 °F)
Sample Inlet Pressure 140 to 275 KPaG (20 to 40 PSIG) to sample panel regulator
Sample Pressure Range 800 to 1400 mbara
Maximum Probe Pressure 275 KPaG (40 PSIG)
Sample Flow Rate Typical 1.0 SLPM (2.1 SCFH)
Recommended Calibration
Two-point calibration in oxygen-free environment (nitrogen) and a second span point
(cylinder gas). Validate with O2 in N2 reference (cylinder gas).
Electrical & Communications
Input Power (Voltage and Max.
Power)
108 to 253 VAC, 50/60 Hz; 5.3W at 120 VAC; 6.6W at 240 VAC or 9 to 30 VDC (CSA), 18 to
30 VDC (IEC/ATEX); 4.7 W at 24 VDC
Communication Analog: Qty 2) 4-20mA sourcing power outputs and (1) 4-20 mA input (sample pressure)
Fieldbus: RS-232C, RS-485, Ethernet 10/100 with Modbus
Output Relays: Qty (2) 250 mA max load (Concentration and Fault Alarms)
USB 2.0 works with Service Software only
4 GB Internal Memory with Internal Data Logging
LCD Display
Concentration, Temperature, Sample Rate, Data Logging, Diagnostics, plus Full Menu for
Setup, Calibration, etc.
Service Software Windows software.
Connected via USB port.
Download data logs, trend and monitor, calibrate and troubleshoot.
Physical
Enclosure Type Type 4X and IP66 rated, 304 and 316 (optional) Stainless Steel
Analyzer Dimensions
280 x 230 x 114 mm (11 x 9 x 4.5 in.) H X W x D (not including Sample Conditioning System)
Controller to Probe Cable Length 0.7 m (2.3 ft.) - Standard
2.5 m (8.2 ft.) and 5.0 m (16.4 ft.) - Optional
Weight 2.2 Kg (4.9 lbs) - analyzer without sample conditioning system
14 kg (31 lbs) - analyzer on a panel
35.4 (78 lbs) - analyzer in enclosure
Sample Probe Construction 316 Stainless Steel
Area Classification - Certification CSA: Class I, Div. 2, Groups A, B, C, and D, T3, NEMA 4X
ATEX/IECEx/UKEX: II 3 G, Ex ec IIC T3 Gc IP66
NOTE: Certification applies to the analyzer only. The enclosure versions of this product are
considered accessories for the product and not included as part of the certification.
Table 35. OXY5500 analyzer specifications