6.2.3 1-Point calibration of an OP-9 sensor
• Precondition: Sensor is connected and put in a low oxygen environment. Sensor constants are already set
correctly (OP-9).
The goal for this example is to calibrate the oxygen sensor.
Step Description Register(s) Value
1 Read the current measurement values. 4899 to 4908 Refer to Table 34.
2
Check to confirm there are no errors, especially error bits 1, 2, 4, 5, and 6. Refer
to Table 8. Proceed only when there are no errors present.
3 Set the calibration values cal0 and T0. 5521 to 5524 1st Float: 66.32
2nd Float: 21.98
Table 33. 1-Point calibration of a OP-9 sensor
Register 4899/4900 Register 4901/4902 Register 4903/4904 Register 4905/4906 Register 4907/4908
Float: Oxygen
Amplitude in mV
Float: Oxygen Phase
shift in degrees
Float: Temperature in °C
Float: Calculated Oxygen
Value in unit
Int32: Error Register.
Refer to Table 8.
50592.62 (Sensor and
environment dependent
value)
66.32 (Sensor and
environment dependent
value)
21.98
This value can be ignored
while calibration process
takes place.
0 (Error code. Should be
0 if a sensor is
connected)
Table 34. Measurement reading for calibration process example