Operating Instructions OXY5500 Gas Analyzer
44 Endress+Hauser
5.13.2 Performing a two-point calibration
To perform a two-point calibration with the connected oxygen sensor, begin by selecting the screens below. When
completed, continue with the procedure detailed in “Calibrating the analyzer” on page 53.
Figure 53. Message window - Sensor type change resets RATA
1. Select Calibration from the SENSOR OPTIONS window. Refer to Figure 54.
Figure 54. Calibration button in the sensor options window
2. Click OK.
A message window will display requesting a response to the following question, “Measurement active. Abort for
Configuration?” Refer to Figure 55.
Figure 55. Message window - Stop measurements during configuration
3. Select Yes to stop the measurement in order to move to the CALIBRATION windows. Use the Up and Down
Arrow buttons to navigate between input fields.